| Division type | Pt100; Pt1000 |
| Temperature coefficient | TCR=3850ppm/K |
| Temperature measuring range | Tolerance class F0.15 (Class A): -50℃~300℃Tolerance class F0.3 (Class B): -70℃~500℃ |
| Temperature error value | Class A: ±(0.15℃+0.002|t|)Class B: ±(0.3℃+0.005|t|)Note: |t| is the absolute value of temperature, unit: ℃ |
| Measuring current | Pt100: 0.3~1.0 mAPt1000: 0.1~0.3 mA (Self‑heating must be considered) |
| Self‑heating | 0.4 K/mW at 0℃ |
| Response time | Water (v=0.4m/s): t0.5 = 0.05s t0.9 = 0.15sAir (v=2m/s): t0.5 = 3.0s t0.9 = 10s"T0.5": Time for 50% step‑change of temperature response"T0.9": Time for 90% step‑change of temperature response |
| Long‑term stability | At 500℃, after 1000 hours, typical drift of R(0℃): 0.04% |
| Temperature measuring environment | Only for dry environment without protection |
| Insulation resistance | >100MΩ at 20℃; >2MΩ at 500℃ |
| Sensor element dimension | 2.3mm2.1mm0.9mm (L×W×H) |
| Lead material | 10 mm platinum‑nickel plated lead |
| Compliance standard | DIN EN60751 |